Abstract: Photoplethysmography is a simple measurement of the
variation in blood volume in tissue. It detects the pulse signal of heart
beat as well as the low frequency signal of vasoconstriction and
vasodilation. The transmission type measurement is limited to only a
few specific positions for example the index finger that have a short
path length for light. The reflectance type measurement can be
conveniently applied on most parts of the body surface. This study
analyzed the factors that determine the quality of reflectance
photoplethysmograph signal including the emitter-detector distance,
wavelength, light intensity, and optical properties of skin tissue.
Light emitting diodes (LEDs) with four different visible
wavelengths were used as the light emitters. A phototransistor was
used as the light detector. A micro translation stage adjusts the
emitter-detector distance from 2 mm to 15 mm.
The reflective photoplethysmograph signals were measured on
different sites. The optimal emitter-detector distance was chosen to
have a large dynamic range for low frequency drifting without signal
saturation and a high perfusion index. Among these four wavelengths,
a yellowish green (571nm) light with a proper emitter-detection
distance of 2mm is the most suitable for obtaining a steady and reliable
reflectance photoplethysmograph signal
Abstract: The process of laser absorption in the skin during
laser irradiation was a critical point in medical application
treatments. Delivery the correct amount of laser light is a critical
element in photodynamic therapy (PDT). More amounts of laser
light able to affect tissues in the skin and small amount not able to
enhance PDT procedure in skin. The knowledge of the skin tone
laser dependent distribution of 635 nm radiation and its penetration
depth in skin is a very important precondition for the investigation of
advantage laser induced effect in (PDT) in epidermis diseases
(psoriasis). The aim of this work was to estimate an optimum effect
of diode laser (635 nm) on the treatment of epidermis diseases in
different color skin. Furthermore, it is to improve safety of laser in
PDT in epidermis diseases treatment. Advanced system analytical
program (ASAP) which is a new approach in investigating the PDT,
dependent on optical properties of different skin color was used in
present work. A two layered Realistic Skin Model (RSM); stratum
corneum and epidermal with red laser (635 nm, 10 mW) were used
for irradiative transfer to study fluence and absorbance in different
penetration for various human skin colors. Several skin tones very
fair, fair, light, medium and dark are used to irradiative transfer. This
investigation involved the principles of laser tissue interaction when
the skin optically injected by a red laser diode. The results
demonstrated that the power characteristic of a laser diode (635 nm)
can affect the treatment of epidermal disease in various color skins.
Power absorption of the various human skins were recorded and
analyzed in order to find the influence of the melanin in PDT
treatment in epidermal disease. A two layered RSM show that the
change in penetration depth in epidermal layer of the color skin has a
larger effect on the distribution of absorbed laser in the skin; this is
due to the variation of the melanin concentration for each color.