Abstract: Photoplethysmography is a simple measurement of the
variation in blood volume in tissue. It detects the pulse signal of heart
beat as well as the low frequency signal of vasoconstriction and
vasodilation. The transmission type measurement is limited to only a
few specific positions for example the index finger that have a short
path length for light. The reflectance type measurement can be
conveniently applied on most parts of the body surface. This study
analyzed the factors that determine the quality of reflectance
photoplethysmograph signal including the emitter-detector distance,
wavelength, light intensity, and optical properties of skin tissue.
Light emitting diodes (LEDs) with four different visible
wavelengths were used as the light emitters. A phototransistor was
used as the light detector. A micro translation stage adjusts the
emitter-detector distance from 2 mm to 15 mm.
The reflective photoplethysmograph signals were measured on
different sites. The optimal emitter-detector distance was chosen to
have a large dynamic range for low frequency drifting without signal
saturation and a high perfusion index. Among these four wavelengths,
a yellowish green (571nm) light with a proper emitter-detection
distance of 2mm is the most suitable for obtaining a steady and reliable
reflectance photoplethysmograph signal
Abstract: Capacitive electrocardiogram (ECG) measurement is an attractive approach for long-term health monitoring. However, there is little literature available on its implementation, especially for multichannel system in standard ECG leads. This paper begins from the design criteria for capacitive ECG measurement and presents a multichannel limb-lead capacitive ECG system with conductive fabric tapes pasted on a double layer PCB as the capacitive sensors. The proposed prototype system incorporates a capacitive driven-body (CDB) circuit to reduce the common-mode power-line interference (PLI). The presented prototype system has been verified to be stable by theoretic analysis and practical long-term experiments. The signal quality is competitive to that acquired by commercial ECG machines. The feasible size and distance of capacitive sensor have also been evaluated by a series of tests. From the test results, it is suggested to be greater than 60 cm2 in sensor size and be smaller than 1.5 mm in distance for capacitive ECG measurement.