Abstract: Artificial joint replacements such as total knee and total hip prosthesis have been applied to the patients who affected by osteoarthritis. Although different material combinations are used for these joints, biopolymers are most commonly preferred materials especially for acetabular cup and tibial component of hip and knee joints respectively. The main limitation that shortens the service life of these prostheses is wear. Wear is complicated phenomena and it must be considered with friction and lubrication. In this study, micro wave (MW) induced argon+oxygen plasma surface modification were applied on ultra-high molecular weight polyethylene (UHMWPE) and vitamin E blended UHMWPE (VE-UHMWPE) biopolymer surfaces to improve surface wettability and wear resistance of the surfaces. Contact angel measurement method was used for determination of wettability. Ball-on-disc wear test was applied under 25% bovine serum lubrication conditions. The results show that surface wettability and wear resistance of both material samples were increased by plasma surface modification.
Abstract: Excessive fretting wear at the taper-trunnion junction
(trunnionosis) apparently contributes to the high failure rates of hip
implants. Implant wear and corrosion lead to the release of metal
particulate debris and subsequent release of metal ions at the tapertrunnion
surface. This results in a type of metal poisoning referred to
as metallosis. The consequences of metal poisoning include;
osteolysis (bone loss), osteoarthritis (pain), aseptic loosening of the
prosthesis and revision surgery. Follow up after revision surgery,
metal debris particles are commonly found in numerous locations. Background: A stable connection between the femoral ball head
(taper) and stem (trunnion) is necessary to prevent relative motions
and corrosion at the taper junction. Hence, the importance of
component assembly cannot be over-emphasized. Therefore, the aim
of this study is to determine the influence of head-stem junction
assembly by press fitting and the subsequent
disengagement/disassembly on the connection strength between the
taper ball head and stem. Methods: CoCr femoral heads were assembled with High stainless
hydrogen steel stem (trunnion) by Push-in i.e. press fit; and
disengaged by pull-out test. The strength and stability of the two
connections were evaluated by measuring the head pull-out forces
according to ISO 7206-10 standards. Findings: The head-stem junction strength linearly increases with
assembly forces.
Abstract: Total hip replacement had been one of the most
successful operations in hip arthritis surgery. The purpose of this
research had been to develop a dynamic hip contact of Thai femoral
bone to analyze the stress distribution on the implant and the strain
distribution on the bone model under daily activities and compared
with the static load simulation. The results showed the different of
maximum von Mises stress 0.14 percent under walking and 0.03
percent under climbing stair condition and the different of equivalent
total strain 0.52 percent under walking and 0.05 percent under
climbing stair condition. The muscular forces should be evaluated
with dynamic condition to reduce the maximum von Mises stress and
equivalent total strain.