Abstract: We present a label-free biosensor based on
electrochemical impedance spectroscopy for the detection of proinflammatory
cytokine Tumor Necrosis Factor (TNF-α). Secretion of
TNF-α has been correlated to the onset of various diseases including
rheumatoid arthritis, Crohn-s disease etc. Gold electrodes were
patterned on a silicon substrate and self assembled monolayer of
dithiobis-succinimidyl propionate was used to develop the biosensor
which achieved a detection limit of ~57fM. A linear relationship was
also observed between increasing TNF-α concentrations and chargetransfer
resistance within a dynamic range of 1pg/ml – 1ng/ml.