Ultra Wideband Breast Cancer Detection by Using SAR for Indication the Tumor Location

This paper presents breast cancer detection by
observing the specific absorption rate (SAR) intensity for
identification tumor location, the tumor is identified in coordinates
(x,y,z) system. We examined the frequency between 4-8 GHz to look
for the most appropriate frequency. Results are simulated in
frequency 4-8 GHz, the model overview include normal breast with
50 mm radian, 5 mm diameter of tumor, and ultra wideband (UWB)
bowtie antenna. The models are created and simulated in CST
Microwave Studio. For this simulation, we changed antenna to 5
location around the breast, the tumor can be detected when an
antenna is close to the tumor location, which the coordinate of
maximum SAR is approximated the tumor location. For reliable, we
experiment by random tumor location to 3 position in the same size
of tumor and simulation the result again by varying the antenna
position in 5 position again, and it also detectable the tumor position
from the antenna that nearby tumor position by maximum value of
SAR, which it can be detected the tumor with precision in all
frequency between 4-8 GHz.





References:
[1] American Cancer Society, "Breast Cancer Fact & Figures 2013-2014,” 2013
[2] M. Klemm, I. J. Craddock, J. A. Leendertz, A. Preece, and R. Benjamin, "Radar-based breast cancer detection using a hemispherical antenna array—experimental results,” IEEE Transactions on Antennas and Propagation, vol. 57, no.6 pp. 1692-1232, 2010.
[3] P. M. Meaney, M. W. Fanning, T. Raynolds, C. J. Fox, Q. Fang, C. A. Kogel, S. P. Popalack, and K. D. Paulsen , "Initial clinical experience with microwave breast imaging in women with normal mammography,” Academic Radiol, vol. 14, pp. 207-218, 2007.
[4] A. Christ, A. Klingenbock, T. Samaras, C. Goiceanu, and N. Kuster, "The dependence of electromagnetic far-field absorption on body tissue composition in the frequency range from 300 MHz to 6 GHz,” IEEE Transaction on Microwave Theory and Techniques, vol. 54 ,no. 5, pp. 2188-2195, 2006.
[5] E. C. Fear, P. M. Meaney, and M. A. Stuchly, "Microwaves for breast cancer detection?,” IEEE Potentials, vol. 22, pp. 12-18, 2003.
[6] P. Thosdee, "Design of an antenna and RF front end transmitter circuit for ultra wideband wireless communication systems,” Suranaree university of technology, 2008.
[7] S. Thanormsuay, P. Thosdee, and C. Thongsopa, "Array of quasi-rhomboid antenna for ultra wideband applications,” ECTI-CON 2008, pp. 293-296, 2008.
[8] N. S. Hassaine, L. Merad, S. M. Meriah, and F. T. Bendimerad, "UWB bowtie slot antenna for breast cancer detection,” World Academy of Science, Engineering and Technology, vol. 6, pp. 1218-1221, 2012.
[9] A. Santorelli, "Breast screening with custom-shaped pulsed microwaves (ch. 4),”McGill University, 2012.
[10] S. A. Winkler, E. Porter, A. Santorelli, M. Coates, and M. Popovic, "Recent progreass in ultra-wideband microwave breast cancer detection,” Ultra-wideband (ICUWB), pp. 182-186, 2012.
[11] S. M. Razavizadeh, "A new link set-up for breast tumor detection,” Power Amplifiers for Wireless and Radio Applications (PAWR), pp. 109-111, 2013.
[12] S. I. Al-Mously, and M. M. Abousetta, "A study of the hand-hold impact on the EM interaction of a cellular handset and a human,”World Academy of Science, Engineering and Technology, Vol. 2, pp. 157-161, 2008.
[13] D. X. Yin, M. Li, and J. L. Li, "Non-invasive breast cancer thermotherapy studies using conformal microstrip antennas,” Antenna, Propagation & EM Theory (ISAPE),pp. 159-162, 2012.