Functional Sample of the Portable Device for Fast Analysis of Explosives

The construction of original functional sample of the portable device for fast analysis of energetic materials has been described in the paper. The portable device consisting of two parts – an original miniaturized microcolumn liquid chromatograph and a unique chemiluminescence detector – has been proposed and realized. In a very short time, this portable device is capable of identifying selectively most of military nitramine- and nitroesterbased explosives as well as inorganic nitrates occurring in trace concentrations in water or in soil. The total time required for the identification of extracts is shorter than 8 minutes.

A Novel Portable Device for Fast Analysis of Energetic Materials in the Environment

Construction of portable device for fast analysis of energetic materials is described in this paper. The developed analytical system consists of two main parts: a miniaturized microcolumn liquid chromatograph of unique construction and original chemiluminescence detector. This novel portable device is able to determine selectively most of nitramine- and nitroester-based explosives as well as inorganic nitrates at trace concentrations in water or soil extracts in less than 8 minutes.