Abstract: The purpose of this study was to investigate the
response of the newly released Gafchromic HD-V2 film for alpha
particle of 5.5 MeV. Gafchromic HD-V2 was exposed to alpha
particles of energy 5 MeV from 241Am for different durations. Then
the films were scanned with a flatbed scanner. The dose response
curve up to 2200 Gy has been achieved. The film’s reproducibility
and sensitivity were evaluated. The results obtained show that the net
optical density increases almost exponentially with the increase in the
exposure time, and it becomes saturated after prolonged exposure
times. The red channel shows the highest sensitivity, with a value of
4 x 10-3 Gy-1 at netOD of 0.4. The inter-film reproducibility was
measured and the relative uncertainty found was 1.7 %, 2.1 % and 2.3
% for grey, red and green channels, respectively.