Abstract: An electronic portal image device (EPID) has become
a method of patient-specific IMRT dose verification for radiotherapy.
Research studies have focused on pre and post-treatment verification,
however, there are currently no interventional procedures using EPID
dosimetry that measure the dose in real time as a mechanism to
ensure that overdoses do not occur and underdoses are detected as
soon as is practically possible. As a result, an EPID-based real time
dose verification system for dynamic IMRT was developed and was
implemented with MATLAB/Simulink. The EPID image acquisition
was set to continuous acquisition mode at 1.4 images per second. The
system defined the time constraint gap, or execution gap at the image
acquisition time, so that every calculation must be completed before
the next image capture is completed. In addition, the