Abstract: Total hip replacement had been one of the most
successful operations in hip arthritis surgery. The purpose of this
research had been to develop a dynamic hip contact of Thai femoral
bone to analyze the stress distribution on the implant and the strain
distribution on the bone model under daily activities and compared
with the static load simulation. The results showed the different of
maximum von Mises stress 0.14 percent under walking and 0.03
percent under climbing stair condition and the different of equivalent
total strain 0.52 percent under walking and 0.05 percent under
climbing stair condition. The muscular forces should be evaluated
with dynamic condition to reduce the maximum von Mises stress and
equivalent total strain.
Abstract: The pullout strength had an effect on the stability of
plate screw fixation when inserted in the cervical spine. Nine
different titanium alloy bone screws were used to test the pullout
strength through finite element analysis. The result showed that the
Moss Miami I can bear the highest pullout force at 1,075 N, which
causes the maximum von Mises stress at 858.87 MPa, a value over
the yield strength of titanium. The bone screw should have large
outer diameter, core diameter and proximal root radius to increase the
pullout strength.