Abstract: A multilayer passive shield composed of low-activity
lead (Pb), copper (Cu), tin (Sn) and iron (Fe) was designed and
manufactured for a coaxial HPGe detector placed at a surface
laboratory for reducing background radiation and radiation dose to
the personnel. The performance of the shield was evaluated and
efficiency curves of the detector were plotted by using of various
standard sources in different distances. Monte Carlo simulations and
a set of TLD chips were used for dose estimation in two distances of
20 and 40 cm. The results show that the shield reduced background
spectrum and the personnel dose more than 95%.