Abstract: In this paper, algorithm estimating the blood pressure
was proposed using the pulse transit time (PTT) as a more convenient
method of measuring the blood pressure. After measuring ECG and
pressure pulse, and photoplethysmography, the PTT was calculated
from the acquired signals. Thereafter, the system to indirectly measure
the systolic pressure and the diastolic pressure was composed using
the statistic method. In comparison between the blood pressure
indirectly measured by proposed algorithm estimating the blood
pressure and real blood pressure measured by conventional
sphygmomanometer, the systolic pressure indicates the mean error of
±3.24mmHg and the standard deviation of 2.53mmHg, while the
diastolic pressure indicates the satisfactory result, that is, the mean
error of ±1.80mmHg and the standard deviation of 1.39mmHg. These
results are satisfied with the regulation of ANSI/AAMI for
certification of sphygmomanometer that real measurement error value
should be within the mean error of ±5mmHg and the standard
deviation of 8mmHg. These results are suggest the possibility of
applying to portable and long time blood pressure monitoring system
hereafter.
Abstract: A analysis on the conventional the blood pressure estimation method using an oscillometric sphygmomanometer was
performed through a computer simulation using an arterial pressure-volume (APV) model. Traditionally, the maximum amplitude algorithm (MAP) was applied on the oscillation waveforms of the APV model to obtain the mean arterial pressure and the characteristic ratio. The estimation of mean arterial pressure and
characteristic ratio was significantly affected with the shape of the blood pressure waveforms and the cutoff frequency of high-pass filter
(HPL) circuitry. Experimental errors are due to these effects when estimating blood pressure. To find out an algorithm independent from
the influence of waveform shapes and parameters of HPL, the volume
oscillation of the APV model and the phase shift of the oscillation with fast fourier transform (FFT) were testified while increasing the cuff
pressure from 1 mmHg to 200 mmHg (1 mmHg per second). The phase shift between the ranges of volume oscillation was then only observed between the systolic and the diastolic blood pressures. The same results were also obtained from the simulations performed on two different the arterial blood pressure waveforms and one
hyperthermia waveform.
Abstract: The linear SEF (Spectral Edge Frequency) parameter
and spectrum analysis method can not reflect the non-linear of EEG.
This method can not contribute to acquire real time analysis and obtain
a high confidence in the clinic due to low discrimination. To solve the
problems, the development of a new index is carried out using the
bispectrum analyzing the EEG(electroencephalogram) including the
non-linear characteristic. After analyzing the bispectrum of the 2
dimension, the most significant power spectrum density peaks appeared abundantly at the specific area in awakening and anesthesia state. These points are utilized to create the new index since many
peaks appeared at the specific area in the frequency coordinate. The measured range of an index was 0-100. An index is 20-50 at an anesthesia, while the index is 90-60 at the awake. New index could afford to effectively discriminate the awake and anesthesia state.
Abstract: The measurement of anesthetic depth is necessary in
anesthesiology. NN10 is very simple method among the RR intervals
analysis methods. NN10 parameter means the numbers of above the 10
ms intervals of the normal to normal RR intervals.
Bispectrum analysis is defined as 2D FFT. EEG signal reflected the
non-linear peristalsis phenomena according to the change brain
function. After analyzing the bispectrum of the 2 dimension, the most
significant power spectrum density peaks appeared abundantly at the
specific area in awakening and anesthesia state. These points are
utilized to create the new index since many peaks appeared at the
specific area in the frequency coordinate. The measured range of an
index was 0-100. An index is 20-50 at an anesthesia, while the index is
90-60 at the awake.
In this paper, the relation between NN10 parameter using ECG and
bisepctrum index using EEG is observed to estimate the depth of
anesthesia during anesthesia and then we estimated the utility of the
anesthetic.
Abstract: The aged are faced with increasing risk for falls. The
aged have the easily fragile bones than others. When falls have
occurred, it is important to detect this emergency state because such
events often lead to more serious illness or even death. A
implementation of PDA system, for detection of emergency situation,
was developed using 3-axis accelerometer in this paper as follows.
The signals were acquired from the 3-axis accelerometer, and then
transmitted to the PDA through Bluetooth module. This system can
classify the human activity, and also detect the emergency state like
falls. When the fall occurs, the system generates the alarm on the
PDA. If a subject does not respond to the alarm, the system determines
whether the current situation is an emergency state or not, and then
sends some information to the emergency center in the case of urgent
situation. Three different studies were conducted on 12 experimental
subjects, with results indicating a good accuracy. The first study was
performed to detect the posture change of human daily activity. The
second study was performed to detect the correct direction of fall. The
third study was conducted to check the classification of the daily
physical activity. Each test was lasted at least 1 min. in third study.
The output of acceleration signal was compared and evaluated by
changing a various posture after attaching a 3-axis accelerometer
module on the chest. The newly developed system has some important
features such as portability, convenience and low cost. One of the
main advantages of this system is that it is available at home
healthcare environment. Another important feature lies in low cost to
manufacture device. The implemented system can detect the fall
accurately, so will be widely used in emergency situation.
Abstract: As the increase of intraoral acidity due to ingestion of sweet foods and acidic beverages usually bring forth a dental caries and a erosion, the measurement of intraoral pH is essential in the study of oral environment. The indwelling intraoral pH telemetry for lasting longer than 24 hours in the mouth was developed to overcome the limits of conventional wire electrode method previously used for salivary and plaque pH measurement, and to assess its effectiveness.