Abstract: In this study three commercial semiconductor devices
were characterized in the laboratory for computed tomography
dosimetry: one photodiode and two phototransistors. It was evaluated
four responses to the irradiation: dose linearity, energy dependence,
angular dependence and loss of sensitivity after X ray exposure. The
results showed that the three devices have proportional response with
the air kerma; the energy dependence displayed for each device
suggests that some calibration factors would be applied for each one;
the angular dependence showed a similar pattern among the three
electronic components. In respect to the fourth parameter analyzed,
one phototransistor has the highest sensitivity however it also showed
the greatest loss of sensitivity with the accumulated dose. The
photodiode was the device with the smaller sensitivity to radiation,
on the other hand, the loss of sensitivity after irradiation is negligible.
Since high accuracy is a desired feature for a dosimeter, the
photodiode can be the most suitable of the three devices for
dosimetry in tomography. The phototransistors can also be used for
CT dosimetry, however it would be necessary a correction factor due
to loss of sensitivity with accumulated dose.
Abstract: Computed tomography (CT) dosimetry normally uses
an ionization chamber 100 mm long to estimate the computed
tomography dose index (CTDI), however some reports have already
indicated that small devices could replace the long ion chamber to
improve quality assurance procedures in CT dosimetry. This paper
presents a novel dosimetry system based in a commercial
phototransistor evaluated for CT dosimetry. Three detector
configurations were developed for this system: with a single, two and
four devices. Dose profile measurements were obtained with them
and their angular response were evaluated. The results showed that
the novel dosimetry system with the phototransistor could be an
alternative for CT dosimetry. It allows to obtain the CT dose profile
in details and also to estimate the CTDI in longer length than the
100 mm pencil chamber. The angular response showed that the one
device detector configuration is the most adequate among the three
configurations analyzed in this study.