Abstract: The paper presents the multi-element synthetic
transmit aperture (MSTA) method with a small number of elements
transmitting and all elements apertures in medical ultrasound
imaging. As compared to the other methods MSTA allows to
increase the system frame rate and provides the best compromise
between penetration depth and lateral resolution.
In the experiments a 128-element linear transducer array with
0.3 mm pitch excited by a burst pulse of 125 ns duration were used.
The comparison of 2D ultrasound images of tissue mimicking
phantom obtained using the STA and the MSTA methods is
presented to demonstrate the benefits of the second approach. The
results were obtained using SA algorithm with transmit and receive
signals correction based on a single element directivity function.
Abstract: The work describes the use of a synthetic transmit
aperture (STA) with a single element transmitting and all elements
receiving in medical ultrasound imaging. STA technique is a novel
approach to today-s commercial systems, where an image is acquired
sequentially one image line at a time that puts a strict limit on the
frame rate and the amount of data needed for high image quality. The
STA imaging allows to acquire data simultaneously from all
directions over a number of emissions, and the full image can be
reconstructed.
In experiments a 32-element linear transducer array with 0.48 mm
inter-element spacing was used. Single element transmission aperture
was used to generate a spherical wave covering the full image region.
The 2D ultrasound images of wire phantom are presented obtained
using the STA and commercial ultrasound scanner Antares to
demonstrate the benefits of the SA imaging.
Abstract: Support Vector Machine (SVM) is a statistical learning tool that was initially developed by Vapnik in 1979 and later developed to a more complex concept of structural risk minimization (SRM). SVM is playing an increasing role in applications to detection problems in various engineering problems, notably in statistical signal processing, pattern recognition, image analysis, and communication systems. In this paper, SVM was applied to the detection of medical ultrasound images in the presence of partially developed speckle noise. The simulation was done for single look and multi-look speckle models to give a complete overlook and insight to the new proposed model of the SVM-based detector. The structure of the SVM was derived and applied to clinical ultrasound images and its performance in terms of the mean square error (MSE) metric was calculated. We showed that the SVM-detected ultrasound images have a very low MSE and are of good quality. The quality of the processed speckled images improved for the multi-look model. Furthermore, the contrast of the SVM detected images was higher than that of the original non-noisy images, indicating that the SVM approach increased the distance between the pixel reflectivity levels (detection hypotheses) in the original images.
Abstract: This paper proposes a method for speckle reduction in
medical ultrasound imaging while preserving the edges with the
added advantages of adaptive noise filtering and speed. A nonlinear
image diffusion method that incorporates local image parameter,
namely, scatterer density in addition to gradient, to weight the
nonlinear diffusion process, is proposed. The method was tested for
the isotropic case with a contrast detail phantom and varieties of
clinical ultrasound images, and then compared to linear and some
other diffusion enhancement methods. Different diffusion parameters
were tested and tuned to best reduce speckle noise and preserve
edges. The method showed superior performance measured both
quantitatively and qualitatively when incorporating scatterer density
into the diffusivity function. The proposed filter can be used as a
preprocessing step for ultrasound image enhancement before
applying automatic segmentation, automatic volumetric calculations,
or 3D ultrasound volume rendering.