High Performance Liquid Chromatography Determination of Urinary Hippuric Acid and Benzoic Acid as Indices for Glue Sniffer Urine

A simple method for the simultaneous determination of hippuric acid and benzoic acid in urine using reversed-phase high performance liquid chromatography was described. Chromatography was performed on a Nova-Pak C18 (3.9 x 150 mm) column with a mobile phase of mixed solution methanol: water: acetic acid (20:80:0.2) and UV detection at 254 nm. The calibration curve was linear within concentration range at 0.125 to 6.0 mg/ml of hippuric acid and benzoic acid. The recovery, accuracy and coefficient variance of hippuric acid were 104.54%, 0.2% and 0.2% respectively and for benzoic acid were 98.48%, 1.25% and 0.60% respectively. The detection limit of this method was 0.01ng/l for hippuric acid and 0.06ng/l for benzoic acid. This method has been applied to the analysis of urine samples from the suspected of toluene abuser or glue sniffer among secondary school students at Johor Bahru.