Abstract: For gamma radiation detection, assemblies having
scintillation crystals and a photomultiplier tube, also there is a
preamplifier connected to the detector because the signals from
photomultiplier tube are of small amplitude. After pre-amplification
the signals are sent to the amplifier and then to the multichannel
analyser. The multichannel analyser sorts all incoming electrical
signals according to their amplitudes and sorts the detected photons
in channels covering small energy intervals. The energy range of
each channel depends on the gain settings of the multichannel
analyser and the high voltage across the photomultiplier tube. The
exit spectrum data of the two main isotopes studied ,putting data in
biomass program ,process it by Matlab program to get the solid
holdup image (solid spherical nuclear fuel)