Abstract: In biomedical implant field, a new formula is given
for the study of Radio Frequency power attenuation by simultaneous
effects of side and angular misalignment of the supply/data transfer
coils. A confrontation with the practical measurements done into a
Faraday cage, allowed a checking of the obtained theoretical results.
The DC supply systems without material connection and the data
transmitters used in the case of biomedical implants, can be well
dimensioned by taking into account the possibility of power
attenuation by misalignment of transfer coils
Abstract: In a transcutanious inductive coupling of a biomedical
implant, a new formula is given for the study of the Radio Frequency
power attenuation by the biological tissue. The loss of the signal
power is related to its interaction with the biological tissue and the
composition of this one. A confrontation with the practical
measurements done with a synthetic muscle into a Faraday cage,
allowed a checking of the obtained theoretical results. The
supply/data transfer systems used in the case of biomedical implants,
can be well dimensioned by taking in account this type of power
attenuation.