Abstract: The paper presents the multi-element synthetic
transmit aperture (MSTA) method with a small number of elements
transmitting and all elements apertures in medical ultrasound
imaging. As compared to the other methods MSTA allows to
increase the system frame rate and provides the best compromise
between penetration depth and lateral resolution.
In the experiments a 128-element linear transducer array with
0.3 mm pitch excited by a burst pulse of 125 ns duration were used.
The comparison of 2D ultrasound images of tissue mimicking
phantom obtained using the STA and the MSTA methods is
presented to demonstrate the benefits of the second approach. The
results were obtained using SA algorithm with transmit and receive
signals correction based on a single element directivity function.
Abstract: The work describes the use of a synthetic transmit
aperture (STA) with a single element transmitting and all elements
receiving in medical ultrasound imaging. STA technique is a novel
approach to today-s commercial systems, where an image is acquired
sequentially one image line at a time that puts a strict limit on the
frame rate and the amount of data needed for high image quality. The
STA imaging allows to acquire data simultaneously from all
directions over a number of emissions, and the full image can be
reconstructed.
In experiments a 32-element linear transducer array with 0.48 mm
inter-element spacing was used. Single element transmission aperture
was used to generate a spherical wave covering the full image region.
The 2D ultrasound images of wire phantom are presented obtained
using the STA and commercial ultrasound scanner Antares to
demonstrate the benefits of the SA imaging.
Abstract: The paper presents the study of synthetic transmit
aperture method applying the Golay coded transmission for medical
ultrasound imaging. Longer coded excitation allows to increase the
total energy of the transmitted signal without increasing the peak
pressure. Signal-to-noise ratio and penetration depth are improved
maintaining high ultrasound image resolution.
In the work the 128-element linear transducer array with 0.3 mm
inter-element spacing excited by one cycle and the 8 and 16-bit
Golay coded sequences at nominal frequencies 4 MHz was used.
Single element transmission aperture was used to generate a spherical
wave covering the full image region and all the elements received the
echo signals. The comparison of 2D ultrasound images of the wire
phantom as well as of the tissue mimicking phantom is presented to
demonstrate the benefits of the coded transmission. The results were
obtained using the synthetic aperture algorithm with transmit and
receive signals correction based on a single element directivity
function.
Abstract: In the paper the study of synthetic transmit aperture
method applying the Golay coded transmission for medical
ultrasound imaging is presented. Longer coded excitation allows to
increase the total energy of the transmitted signal without increasing
the peak pressure. Moreover signal-to-noise ratio and penetration
depth are improved while maintaining high ultrasound image
resolution. In the work the 128-element linear transducer array with
0.3 mm inter-element spacing excited by one cycle and the 8 and 16-
bit Golay coded sequences at nominal frequency 4 MHz was used. To
generate a spherical wave covering the full image region a single
element transmission aperture was used and all the elements received
the echo signals. The comparison of 2D ultrasound images of the
tissue mimicking phantom and in vitro measurements of the beef liver
is presented to illustrate the benefits of the coded transmission. The
results were obtained using the synthetic aperture algorithm with
transmit and receive signals correction based on a single element
directivity function.
Abstract: The paper presents the optimization problem for the
multi-element synthetic transmit aperture method (MSTA) in
ultrasound imaging applications. The optimal choice of the transmit
aperture size is performed as a trade-off between the lateral
resolution, penetration depth and the frame rate. Results of the
analysis obtained by a developed optimization algorithm are
presented. Maximum penetration depth and the best lateral resolution
at given depths are chosen as the optimization criteria. The
optimization algorithm was tested using synthetic aperture data of
point reflectors simulated by Filed II program for MatlabĀ® for the
case of 5MHz 128-element linear transducer array with 0.48 mm
pitch are presented. The visualization of experimentally obtained
synthetic aperture data of a tissue mimicking phantom and in vitro
measurements of the beef liver are also shown. The data were
obtained using the SonixTOUCH Research systemequipped with a
linear 4MHz 128 element transducerwith 0.3 mm element pitch, 0.28
mm element width and 70% fractional bandwidth was excited by one
sine cycle pulse burst of transducer's center frequency.